UV Visible Spectrophotometer

Order Code: 21224902.1.2

Category: General Lab Equipment II

Purpose: For educational purpose to train technologists and engineers on how to operate UV-Vis equipment to analyze the chemical structure of substances in water and sediment samples. System Design: Working Mode: PC controlled, scanning type ...



SPECIFICATION

Purpose: For educational purpose to train technologists and engineers on how to operate UV-Vis equipment to analyze the chemical structure of substances in water and sediment samples.

System Design: Working Mode: PC controlled, scanning type

  • Optical System

  • Sealed optical system against dust & vapor

  • Reflecting mirrors coated with quartz

  • Double Beam with samples and reference cell holder

  • Monochromator :

  • Czerny Turner mounting

  • Grating should be : 1200 lines/mm or better

  • Wavelength Range :

  • Range : 190nm to 1100nm or better Resolution : 0.1nm

  • Spectral Band Width : Band width: 0.20–4.00 nm, Variable width 0.1 nm steps

  • Scanning Speed :

  • Speed: 10,000 nm/min or more Scan interval: 0.1nm ? Wavelength accuracy ± 0.03 nm or better

  • Wavelength reproducibility: ± 0.01nm or better

  • Photometric Linear Range 4 Abs or better

  • Photometric Stability <0.0004 Abs or better

  • Photometric noise ≤0.0001 Abs or better

  • Sample compartment : Shall support one for reference cell holder &

  • multi cell sample holders / Cuvette /

  • Sample compartment access Top & front

PC / Software / Data processing:

  • 3. PC with printer should be supplied along with the instrument.

  • Supplied PC compatible for operation & monitoring of the instrument.

  • High resolution LCD color display

  • 4. Software related

  • Licensed software should be provided, Microsoft Windows operating platform with easy user interface

  • Basic software for device control & data analysis.

  • Easy and user adapted data-base transfer to external software for further processing.

  • Data storage to be provided for all data.

  • Instrument software shall include

  • Self diagnostic feature.

  • Software shall include quality control protocols as well.

Accessories:

  • The UV-Vis should be provided with necessary accessories. The presence of these accessories should not limit or restrict our usage as independent equipment.

  • Example

  • Branded PC – with preloaded window

  • Printer

  • Color LCD monitor

  • Diffuse reflectance Accessories, Specular reflectance accessories, Reference Standards, Magnetic Stirrer, Miniature Stirrer systems

  • Not less than 6 sets of quartz cuvette to be offered

  • Offered accessories to provide extra capabilities

Consumables & Spare parts:

  • The UV-Vis should be provided with necessary consumables. The presence of these consumables should not limit or restrict our usage as independent equipment.

  • Start up kit of Glass wares & Consumables for two years operation as recommended. A list of spares included with the equipment, sufficient number of rapidly wearing & consumable parts are to be included to cover the guarantee period

  • Example :- Absorbance / Transmission / Calibration Filters, UV- VIS-Cuvette / Cell all types, D2 & Visible lamps compatible to the equipment, Considering expiry date.

Power requirement:

  • 220V, 50Hz for single phase,
  • IEC World plug type F or EEC3/7 (Schuko)

 

Warranty: 2 years starting from the date of acceptance testing approval

Installation and commissioning: The equipment shall be installed, configured, and thoroughly tested by the manufacturer before handing over.

Training: Training of users in operation and basic maintenance are to be provided for 3 days, 8 hours a day

Technical Documentation: User, installation and maintenance manuals shall be supplied in English language

Enquiry Form