Order Code: 21224973.18
Category: General Lab Equipment II
Function : Measurement of 2/3/4-pole earth resistance, soil resistivity, earth voltage, AC voltage Measurement Range : Earth Resistance: 0.00?-30.00 k?
Function : Measurement of 2/3/4-pole earth resistance, soil resistivity, earth voltage, AC voltage
Measurement Range : Earth Resistance: 0.00Ω-30.00 kΩ
: Soil Resistivity: 0.00Ω-9000kΩ
Measuring Mode : Precise 4-pole measurement, 3-pole measurement, simple 2-pole measurement
Measuring Method : Earth Resistance: rated current change-pole method, measurement current 20mA Max
: Soil Resistivity: 4-pole measurement
: Earth Voltage: average rectification(between P(S)-ES)
Test Frequency : 128Hz/111Hz/105Hz/94Hz(AFC)
Short-circuit Test Current : AC 20mA max
Open-circuit Test Current : AC 40V max
Test Voltage Wave : Sine wave
Electrode Distance Range: Can be set 1m-100m
Shift : Earth Resistance: 0.00Ω-30.00 kΩ, automatic shift
: Soil Resistivity: 0.00Ω-9000kΩ, automatic shift
Backlight : Blue screen backlight, suitable for dim places
Display Mode : 4-digital super-large LCD display, blue screen backlight
Measuring Indicator : During measurement, LED flash indicator, LCD count down display, progress bar indicator
LCD Frame Dimension : 128mm×75mm
LCD Window Dimension : 124mm×67mm
Standard Test Wire : 4 wires: each for red 20m, black 20m, yellow 10m, and green 10m
Simple Test Wire : 2 wires: each for red 1.6m and black 1.6m
Auxiliary Earthing Rod : 4 rods: Φ10mm×150mm
Measuring Rate : Earth voltage: about 3 times/second
Earth resistance, soil resistivity: about 5 seconds/time
Line Voltage : below AC 600V
Measuring Times : Over 5000 times (Short-circuit test, interval time should be at least 30 seconds)
RS232 Interface : RS232 interface, software supervision, storage data can be uploaded to computer, saved or printed.
Communication Wire : One piece of RS232 communication wire, with length 1.5m
Data Storage : 300 sets, “MEM” icon storage indicator, flash display “FULL” icon to indicate storage is full
Data Hold : Data hold function: “HOLD” icon display
Data Access : Data read function: “READ” icon display
Overflow Display : Exceeding measuring range overflow function: “OL” icon display
Interference Test : Recognize interference signal automatically, “NOISE” icon display when interference voltage exceed 5V
Auxiliary Earthing Test : Can measure auxiliary earth resistance, 0.00KΩ-30kΩ(100 R+rC<50kΩ, 100 R+rP<50kΩ)
Operating Temperature & Humidity : -10? -40? , below 80%rh
Storage temperature & humidity : -20? -60? , below 70%rh
Alarm Function : When measuring value exceeds alarm setting value, there is alarm Sound
Battery Voltage : When battery voltage decreases to about 7.5V, battery voltage low icon will display, reminding to replace battery.
Overload Protection : Measuring earth resistance: between each interfaces of C(H)-E?P(S)-ES, AC 280V/3 seconds
Insulation Resistance : Over 20MΩ (between circuit and enclosure it is 500V)
Withstanding Voltage : AC 3700V/rms (Between circuit and enclosure)
Electromagnetic Features : IEC61326(EMC)
Protection Type : IEC61010-1 (CAT ? 300 V?CAT IV 150V?Pollution 2), IEC61010-031, IEC61557-1(Earth resistance), IEC61557-5 (Soil resistivity), JJG 366-2004
Power Consumption : Standby: about 20mA (Backlight shut off)
: Boot and with backlight: about 45mA (25mA without backlight)
: Measurement: about 100mA (Backlight shut off)
Power Supply : DC 9V(Zi-Mn dry battery R14S 1.5V 6 PCS, continuous standby for 300 hours )
Accessories : Standard testing wire 4nos., Simple testing wire 2nos., earth rods 4pcs., Battery, Manual Software CD, RS232 Communication cable & Carrying case.