Single Phase Mobile Meter Test Bench

Order Code: 23246531.1

Category: General Lab Equipment II

Environmental Parameters  Ambient Temperature : -10°C to +40°C Relative Humidity : 35% - 85% Accuracy Class : 0.05% , 0.1% Power Supply : One Phase AC 100-265V ; Frequency 50/60 Hz Test to be Performed  Pre-wa...



SPECIFICATION

Environmental Parameters 

  • Ambient Temperature : -10°C to +40°C
  • Relative Humidity : 35% - 85%
  • Accuracy Class : 0.05% , 0.1%
  • Power Supply : One Phase AC 100-265V ; Frequency 50/60 Hz

Test to be Performed 

  • Pre-warming, Accuracy test per IS/IEC, Starting current test, Creep test, Dial test, Local checking the testing results, esting 3 pcs Or 2pcs meter synchronously

Construction and Component of System

  • Microprocessor-based, modular type of 19 rack design, in built voltage amplifier and current amplifier

AC Voltage Output

  • Voltage Range - 57.7V, 100V, 220V, 380V ; max 500V

AC Current Output 

  • Current Range - 200ma, 1A, 5A, 20A, 100A ; max 120A

Power Output 

  • Active Power Output Stability < 0.01% RG120s 
  • Reactive Power Output Stability < 0.02% RG120s

Phase Output 

  • Output Adjustment Range 0° - 359.999°
  • Output Adjustment Fineness 10, 1, 0.1, as optional

Power Factor 

  • Adjustment Range : -1~0~1

Frequency Output 

  • Adjustment Range - 40Hz to 70Hz

Voltage/Current/Harmonic Setting 

  • Harmonic Number 2-51 times
  • Harmonic Content 0-40 %

Power Energy Measurement Error 

  • Active Power Energy 0.05%RG 
  • Reactive Power Energy 0.1%RG

Power Pulse Output 

  • Power pulse type active and reactive pulse
  • Active power pulse output 5V, 10mA
  • Power Pulse Input 
  • Support active and reactive pulse,
  • The highest frequency power pulse input is 180K

Mechanical Parameters 

  • Dimensions (WxDxH) : 455 x 440 x 150 mm
  • Weight (Kg) : 12 Kg

Enquiry Form