Product Details:
We are offering high quality "Multi Parameter Monitor" 12 inch Five Standard parameters: ECG, NIBP, SPO2, PR / HR
Technical specification:
ECG:
Lead Mode: 5 Leads ( R, L, F, N, C or RA, LA, LL, RL,V)Lead selection: I, II, III, avR, avL, avF, V, Waveform: 2 ch Lead mode: 3 Leads (R, L, F or RA, LA, LL) Lead selection: I, II, III, Waveform: 1 chGain: ?2.5mm/mV, ?5.0mm/mV, ?10mm/mV, ?20mm/mV, auto
HR and Alarm:
- Range Adult : 15 ~ 300 bpm
- Neo/Ped : 15 ~ 350 bpm
- Accuracy : ± 1% or ± 1bpm, which great
- Resolution : 1 bpm
- Sensitivity : > 200 (uV P-P)
- Differential Input Impedance : > 5 MΩ
- Monitor : > 105 dB
- Operation : > 105 dB
- Diagnosis : > 85 dB
- Electrode offset potential : ±300mV
- Leakage Current : < 10 uA
- Baseline Recovery : < 3 S After Defi.
- ECG Signal Range : ±8 m V ( Vp-p )
- Surgery : 1 ~ 15 Hz
- Monitor : 0.5 ~ 35 Hz
- Diagnostic : 0.05 ~ 100 Hz
- Calibration Signal : 1 (mV p-p), Accuracy : ±5%
- ST Segment Monitoring Range : Measure and Alarm -2.0 ~ +2.0 mV
- ARR Detecting
- Type : ASYSTOLE, VFIB/VTAC, COUPLET, BIGEMINY, TRIGEMINY, R ON T, VT>2, PVC, TACHY, BRADY,
SpO2:
- Measuring Range: 0 ~ 100 %
- Alarm Range: 0 ~ 100
- Resolution: 1 %
- Accuracy : 70% ~ 100% ±2 %, 0% ~ 69% unspecified
- Actualization interval: about 1 Sec.
- Alarm Delay: 10 Sec
NIBP:
- Method: Oscillometric
- Mode: Manual, Auto, STAT
- Measuring Period in STAT Mode: 5 Min
- Pulse Rate Range: 40 ~ 240 bpm
- Alarm Type: SYS, DIA, MEAN
- Measuring and alarm range
- SYS: 40 ~ 270 mmHg
- DIA: 10 ~ 215 mmHg
- MEAN: 20 ~ 235 mmHg
- Pediatric Mode
- SYS: 40 ~ 200 mmHg
- DIA: 10 ~ 150 mmHg
- MEAN:20 ~ 165 mmHg
- Neonatal Mode
- SYS: 40 ~ 135 mmHg
- DIA:10 ~ 100 mmHg
- MEAN:20 ~ 110 mmHg
- Resolution Pressure:1mmHg
- Accuracy Pressure Maximum Mean error: ±5mmHg
- Maximum Standard deviation: ±8mmHg
- Adult Mode: 297±3 mmHg
- Pediatric Mode: 240±3 mmHg
- Neonatal Mode: 147±3 mmHg