Oxidation Stability Apparatus

Order Code: 24258380.1

Category: General Lab Equipment V

Technical Specifications for Oxidation Stability Test Apparatus Test Standards 1. Oxidation Bomb Material: Stainless steel (SS), manufactured to standard dimensions. Safety Pressure Rating: Must withstand a pressure ...



SPECIFICATION

Technical Specifications for Oxidation Stability Test Apparatus

Test Standards

  • 1. Oxidation Bomb

  • Material: Stainless steel (SS), manufactured to standard dimensions.

  • Safety Pressure Rating: Must withstand a pressure of 1241 kPa at 99°C.

  • Sealing System: Includes PTFE gaskets, securing screws, and a wrench for tight closure.

  • Pressure Relief System: Integrated with oxygen manifold for controlled release.

2. Pressure Measurement System

  • Measurement Range: 0–1100 kPa, with graduation in 1 psi intervals.

  • Accuracy: Better than ±0.5 psi.

  • Functionality:

    • Real-time electronic pressure measurement and logging.

    • Capable of recording pressure vs. time with high accuracy.

3. Oil Bath System

  • Temperature Control: Maintains temperature at 99°C ± 0.5°C.

  • Gradient Control: Temperature gradient not exceeding 0.5°C across the bath.

  • Heat Recovery: Sufficient thermal capacity to recover to set temperature within 60 minutes after bomb immersion.

  • Submersion Level: Bomb top must remain at least 50 mm below the oil surface.

  • Capacity: Must accommodate four (4) oxidation bombs simultaneously.

  • Temperature Range: From ambient to minimum 105°C.

  • Bath Medium: Compatible with standard test oils.

  • Power Supply: 220 V, 50 Hz.

4. General Requirements

  • Plug-and-Play Operation: Equipment should be ready for immediate use upon installation.

  • Warranty: Unconditional 12-month warranty from date of installation.

  • Spare Parts: Supply of minimum required spares for trouble-free operation.

  • Documentation:

    • Hard and soft copies of operating instructions and service manuals.

    • Must include complete schematic diagrams of the system.

5. Spare / Additional Requirements

  • Thermometer: Range 95–103°C, accuracy ±0.1°C.

  • Pressure Line: Equipped with a quick-release coupling for the bomb’s needle valve and threaded fitting for oxygen cylinder connection.

  • Glass Sample Dishes: Included.

  • Accessories & Spares List:

    • Please provide a complete list of all essential accessories and spare parts.

    • Availability of these items should be ensured for at least 5 years from the date of equipment supply.

6. Installation, Commissioning, and Training

  • Kindly confirm:

    • Charges for installation and commissioning of the equipment.

    • Charges for 2-day training for staff.

 

Enquiry Form