Mechanical Ventilator

Order Code: 25268424.25

Category: General Lab Equipment V

Common Features Operating Modes: Pneumatically driven & electronically controlled; supports closed, semi-closed, and semi-open systems. Clinical Utility: Ideal for emergency resuscitation and respiratory therapy. Display:...



SPECIFICATION

Common Features

  • Operating Modes: Pneumatically driven & electronically controlled; supports closed, semi-closed, and semi-open systems.

  • Clinical Utility: Ideal for emergency resuscitation and respiratory therapy.

  • Display: Built-in 10.4-inch high-visibility color TFT display capable of showing:

    • Pressure-Time, Frequency-Time, Pressure-Volume, Frequency-Volume curves

    • Any two waveforms simultaneously

  • Trigger Options: Pressure trigger, Flow trigger

  • Sigh Function: Inserts 1–8 sigh breaths per 100 breaths, each ≥1.5 times tidal volume

  • Flow Sensors: Dual electronic flow sensors & dual respiratory modules for precision

  • Backup Functionality: Auto-switches to standby mode during AC power failure

  • Sensor System: Built-in infrared turbine flow sensor—stable and accurate

  • Humidifier: 9 levels adjustable for patient comfort

  • Structure: Eco-friendly, corrosion-resistant, and equipped with automated startup self-check & self-calibration

Monitoring Parameters

  • Tidal Volume

  • Respiratory Rate

  • Airway Pressure

  • I:E Ratio

  • Inspiratory & Expiratory Phase Status

  • Trigger Monitoring

Alarm Systems

  • Visual & Audible Alarms for:

    • Ventilation volume

    • Tidal volume

    • Oxygen concentration

    • Asphyxia

    • Airway pressure

    • Intubation disconnection

    • Low oxygen

    • Power failure

  • Alarm History Query Function


Advanced Model Exclusive

  • Integrated medical-grade air compressor for continuous, quiet operation

  • Output Gas Pressure: 0.4MPa ±20%

  • Continuous Output Flow: ≥20 L/min

  • Instantaneous Flow: ≥60 L/min

Ventilation Modes

  • A/C (Assist Control)

  • IPPV, SIPPV, SIMV, PSIMV

  • PSV, CPAP

  • PCV (Pressure-Controlled Ventilation)

  • VCV (Volume-Controlled Ventilation)

  • SPONT (Spontaneous)

  • MANUAL

Technical Specifications

Parameter Range / Value
Display 10.4-inch color TFT
Minute Volume ≥18 L/min
Tidal Volume (Adjustable) 20–1500 ml (Display: 0–2000 ml)
Respiratory Rate 1–99 bpm
I:E Ratio 4:1 to 1:4
PEEP 0–10 cmH?O
SIMV Rate 1–20 bpm
Pressure Trigger -10 to +10 cmH?O
Flow Trigger Sensitivity 1–10 L/min
Inspiratory Plateau 0–50% of inspiratory time
O? Concentration 21% – 100% adjustable
Power Supply (Main) 220V AC ±10%, 50 Hz, 700VA
Backup Battery ≥ 4 hours
Gas Supply 280–600 kPa (medical O? & compressed air)
Pressure Limit 1.0–6.0 kPa (Max Operating ≤6.0 kPa)
Tidal Volume Alarm Upper: 10–2000 ml, Lower: 0–1800 ml
Airway Pressure Alarm Upper: 0.1–6 kPa, Lower: 0–5 kPa
Oxygen Alarm Range Upper: 22–100%, Lower: 21–80%

 