Intrinsically Safe Multimeters

Order Code: 23246498.3

Category: General Lab Equipment III

The Digital Multimeter shall be ATEX compliant (explosion proof) and IS (Intrinsically Safe) True RMS Digital Multimeter Fluke 28 II Ex complete with carrying case, operations/service manuals and electronic combo test leads kit for use in hazardous a...



SPECIFICATION

The Digital Multimeter shall be ATEX compliant (explosion proof) and IS (Intrinsically Safe) True RMS Digital Multimeter Fluke 28 II Ex complete with carrying case, operations/service manuals and electronic combo test leads kit for use in hazardous areas with the technical specifications given below. 
Components
a) True RMS EXII IS a digital Multimeter,
b) Operation and Service Manuals
c) Calibration certificate
d) Compact lightweight soft case.
e) Industrial strength test lead probe kit:-
• Grip Silicone Test Lead Set, one pair (red, black) >1.5m test leads with safety-shrouded std. diameter banana plugs. 
• Electronic Test Probe Set, one pair (red, black) of very small 1mm tips to access hard to hard-to-reach test points with replaceable steel and spring-loaded gold tips.
• Grip Hook Clip Set, one pair (red, black) profile narrows to 5.6mm at tip, hook opening 6.4mm at the front, 2mm at the base with
• Integrated DMM Temperature Probe, one pair (red, black) Type-K thermocouple with standard banana jack 
No. Parameter Value
1)  Functions Measured: AC/DC Current, AC/DC Voltage, Capacitance, Diode, Duty Cycle, Frequency, Resistance, Conductance, Temperature
2)  True RMS: Yes
3)  Voltage Measurement Range: 0.1 mV to 1000 V
4)  Voltage Measurement Resolution: 0.1 mV
5)  DC Voltage Measurement Accuracy: ±0.05% + 1digit
6)  AC Voltage Measurement Accuracy: ±0.7% +4digit
7)  Current Measurement Range: 0.1 μA to 20 A
8)  Current Measurement Resolution: 0.01A 
9)  DC Current Measurement Accuracy: ±0.2% + 4digits
10)  AC Current Measurement Accuracy: ±1.0% + 2digits
11)  Resistance Measurement Range: 0.1 Ω to 50 MΩ 
12)  Resistance Measurement Resolution: 0.1Ω
13)  Resistance Measurement Accuracy: ±0.2% + 1digit
14)  Conductance Measurement Range: 60.00 nS 
15)  Conductance Measurement Accuracy: ±1.0% + 10digits
16)  Diode Test Measurement Range: 2.0 V
17)  Diode Test Measurement Accuracy: ±2.0% + 1digit
18)  Duty Cycle Measurement Range: 0.0% to 99.9%
19)  Duty Cycle Measurement Accuracy: ±(0.2% per kHz + 0.1%) for rise times <1 µs
20)  Conductance Measurement Range: 60.00 nS 
21)  Conductance Measurement Accuracy: ±1.0% + 10digits
22)  Capacitance Measurement Range: 10 nF to 9999 μF
23)  Capacitance Measurement Accuracy: ±1.0% + 2digits
24)  Measurement Frequency range: 0.5 Hz to 199.99 kHz
25)  Frequency Measurement Accuracy: ±0.005% + 1digit
26)  Frequency Measurement resolution: 0.01 kHz
27)  Temperature Measurement range: -200°C to +1090°C 
28)  Temperature Measurement Accuracy: ±1.0% + 10digits
29)  Temperature Measurement Resolution: 0.1°C
30)  Battery Life: > 400 hours
31)  Mechanical and General Specifications: Handheld, Lightweight < 700g 63.5 x 100 x 198.1mm
32)  Display Type: LCD with backlight
33)  Shock Resistance: Tested for drops > 3 meters
34)  Safety Certification   ATEX 
Over Voltage Protection
Intrinsic Safety II 2 G Ex ia IIC T4 Gb 
II 2 D Ex ia IIIC T130°C Db 
I M1 Ex ia I Ma
EN 61010–1 to 1000V CAT III
EN 61010–1 to 600V CAT IV
1 to 600V CAT IV IP rating IP67
35)  External Protection: RED Rubber holster
36)  Calibration Certificate: UKAS or RS or internationally recognized accredited calibration body
37)  Warranty: Three Years

Enquiry Form