Hydride Generator

Order Code: 19204625.1

Category: General Lab Equipment III

Hydride generator should be highly sensitive   Precision should be 1-2 % RSD at ppb levels and results should achieve using integration of the steady state signal for a few seconds   Flow Injection should produce transient ...



SPECIFICATION

  • Hydride generator should be highly sensitive 
  •  Precision should be 1-2 % RSD at ppb levels and results should achieve using integration of the steady state signal for a few seconds 
  •  Flow Injection should produce transient signals, requiring separate injections for each replicate reading. Each replicate should consume several milliliters of sample to ensure the injection loop is filled 
  •  Sample throughput should be high and results should be achieved in less than 1 minute allowing determination of up to 70 samples/hour 
  •  The integral pump should provide continuous flow conditions with automatic reagent addition 
  •  Stability should be high and the continuous flow conditions should ensure a stable baseline and eliminate varying blank levels 
  •  Sample consumption should be low: < 8 mL per element during analysis 
  •  Cross-contamination should be eliminated - modules should be simply exchanged when switching between elements with conflicting chemistries 
  •  Simple and automatic operation - each sample should be automatically combined with a continuous flow of acid and reagent. 

Enquiry Form