Order Code: 23246880.3
Category: General Lab Equipment II
AC Power?220V±10%, 50 HZ, 20 VA,-Battery Life?5000V@100M,- Range accuracy: 5000V-Unspecified: < 100k,- 5%: 100k-100G- 20%: 100G-1T-Unspecified: > 1T- OFF: Power Off (No operation, 3 minutes automatic shutdown) H.V: Set test voltage:...
Instrument using:
1. Select more than 1 minute length measurement.
2. Select the appropriate voltage and start.
3. The instrument automatically record 15 s and 60 s resistance, and calculate the absorption than after 60 seconds.
measured PI :
When measuring high insulation resistance, the sample should be in the position a conductor surface protection between the two measuring end ring, and the conductor protection ring is connected to the instrument with a test line G side, try to eliminate the measuring error caused by the surface leakage current, security testing is accurate.
Required Document: