High Voltage Insulation Testers

Order Code: 22236247.6

Category: General Lab Equipment II

Specification AC Power?220V±10%, 50 HZ, 20 VA Battery Life?5000V@100M Range accuracy: 5000V Unspecified: <100k 5%: 100k-100G 20%: 100G-1T Unspecified: > 1T OFF: Power Off  (No operation, 3 minutes automatic ...



SPECIFICATION

Specification

  • AC Power?220V±10%, 50 HZ, 20 VA
  • Battery Life?5000V@100M
  • Range accuracy: 5000V
  • Unspecified: <100k
  • 5%: 100k-100G
  • 20%: 100G-1T
  • Unspecified: > 1T
  • OFF: Power Off  (No operation, 3 minutes automatic shutdown)
  • H.V: Set test voltage: 500V?1000V?2500V?5000V
  • Set The test time: 01:00>10:00>30:00

Set The Filter

  • 0?without filtering; (no interference case recommended)
  • 1?hardware low-pass filter; (applicable to most situations)
  • 2?software Low pass filter;High resistance (greater than 100G recommended choose)
  • 3?hardware + software low-pass filter
  • (high resistance is greater than 1T recommend choose)
     

Instrument using

  • Select more than 1 minute length measurement
  • Select the appropriate voltage and start
  • The instrument automatically record 15 s and 60 s resistance, and calculate the absorption than after 60 seconds
  • Absorption ratio calculation method: DAR = R60 / R15

Measured PI 

  • Select more than 10 minute length measurement
  • Select the appropriate voltage and start
  • The instrument automatically record 15S, 60S and 600S resistance, and calculate the polarization index after 600 seconds
  • Polarization index calculation method PI= R600/R60

Test Connection Points for Attention

  • Tested to confirm product safety grounding, try not charged
  • Confirm the instrument E end (ground) has been grounded
  • The use of G terminal (protection ring) (this machine is of low voltage side of the screen)
  • When measuring high insulation resistance, the sample should be in the position a conductor surface protection between the two measuring end ring, and the conductor protection ring is connected to the instrument with a test line G side, try to eliminate the measuring error caused by the surface leakage current, security testing is accurate

Documents

  • Manufacturers Authorization Letter Submitted
  • ISO Certificate Submitted
  • Specific Product CE Certificate Submitted
  • Catalog Submitted
  • Warranty: 01 year warranty with service

Enquiry Form