High Pressure Autoclave (Steam sterilization)

Order Code: 22235922.57

Category: General Lab Equipment I

Vertical autoclave, universal basic version for microbiological standard laboratory to sterilize liquids, instruments, glassware, plastic articles or general infectious waste. Triple walled construction; chamber, basket, door lid, doorframe...



SPECIFICATION

  • Vertical autoclave, universal basic version for microbiological standard laboratory to sterilize liquids, instruments, glassware, plastic articles or general infectious waste.
  • Triple walled construction; chamber, basket, door lid, doorframe, bolts made of corrosion-resistant material and able to prevent stress cracking preferably made of high grade stainless steel sheet of SS-304 grade. Housing with SS legs
  • Pressure vessel should be Hydraulic tested at factory with minimum Hydrostatic Pressure: 2.5 kg/cm sq. (35 psi)
  • Working Chamber volume: approx. at lest 150 liters.
  • Electrically heated by immersion type heaters bearing ISI mark.
  • Fast safety lid lock with silicone gasket, it may be radial locking, automatic locking, single lever locking, fly nut assembly mechanism and with heat resistant/safety handle.
  • Manual water feed system with water level indicator, pressure gauge, steam release cock, spring loaded safety valve, water inlet and water valves
  • Accessories to be supplied include stainless steel basket (where 2 fit in autoclave directly plus two spare total 4), stainless steel wire basket (where 2 fit in autoclave directly plus two spare total 4), Chemical indicator tape for sterilization (2), Biological indicator (100), spare heating elements (two), fuses (10) and silicone gaskets (2).
  • Automatic Water Cut-off Device – To protect the heaters from running dry and to ensure that the machine is automatically switched off in case the desired water level falls below the prescribed level
  • Working temperature: 121°C, Maximum operating temperature: 134 °C (273 °F).
  • Working pressure: 15 PSI, Maximum operating pressure: 2.5 bar or 36 PSI
  • Timer with Alarm System - To regulate the sterilization time of the media to be sterilized with a buzzer, Sterilization timer: 1–99 minutes.
  • A visual chamber gauge, which easily identifies pressure in the chamber, must be accessible to the operator as a backup for reading pressure gauge when no electrical power is available.
  • Micro-processor temperature control system with sensor-with user changeable set temperature and timer option. The microprocessor controls the desired temperature (pressure automatically regulated) by cutting off the current to the heating element automatically & restart the  Mechanism as required. The control panel to be mounted so that the components sensitive to steam and heat are protected. Large LCD display showing:
  • Chamber Temperature
  • Sterilization time
  • Alarm information.
  • Alarm: audible, with display on dysfunction & after completion of sterilization cycle.
  • Electrical control box, fitted with toggle switch, indicating Neon lamps for Autoclave ON/OFF status, heater ON/Off status.
  • Over-temperature and over-pressure protection limiter
  • Electrical requirements: Equipment to work on 230 ± 10 volts single phase, 50 Hz, plug type adopted to local country scenario, Voltage regulator of appropriate rating to be included to cope with 160-260 V. The line cord / Power cord supplied with the equipment shall be of acceptable durability, length, and current carrying capacity complying
     