Order Code: 21224761.56
Category: General Lab Equipment III
(1) Hall effect measurement system (HMS) should have automated magnet movement, variable temperature capability (2) The working principle should be the Van der Pauw method (2) Temperature range from 80 de...
(1) Hall effect measurement system (HMS) should have automated magnet movement, variable temperature capability |
(2) The working principle should be the Van der Pauw method |
(2) Temperature range from 80 deg.K to 350 deg.K |
(3) Should have analysis software running on MS Windows |
(4) Automated system should work without user intervention |
(5) The system should ramp up user-defined temperature, then stabilize and make measurements and plot various temperature-dependent material properties I |
(6)Should have the following test |
(6.1) Temperature versus carrier density |
(6.2) Temperature Versus Mobility |
(6.3) Temperature Vs resistivity |
(6.4) Temperature Vs Hall coefficient |
(6.5) Temperature Vs conductivity |
(7)Sample size 5 by 5 MM up to 15 by 15 MM |
(8)Minimum Hall voltage 1 microvolt |
(9)Current source 1nA - 20mA |
(10)Permanent magnet 30-40 MM dia and 0.5-0.6T flux density |
(11)Optional room temperature/LN2 temperature sample kit lid with a sample board should be included with a sample board should be included |
(12)Software installed 17, 8GEN laptop should be provided |
with 32GB RAM, 1TSSD, 1T HDD configuration |
(13)Installation and a minimum 3 days of training should begin |
(14) A Detailed datasheet from the manufacturer should be provided with the offer I |