Purpose: For educational purpose to train technologists and engineers on how to operate GC to make analysis of volatile samples with GC. System Design: System must be capable of effluent splitting, back flushing & column switching/changing wit...
Purpose: For educational purpose to train technologists and engineers on how to operate GC to make analysis of volatile samples with GC.
System Design: System must be capable of effluent splitting, back flushing & column switching/changing without breaking of vacuum along with software controlled splitter to acquire chromatograms for at least three detectors simultaneously with more than two inlets.
Chromatographic Performance:
Retention time performance
Retention time repeatability - < 0.0008 min
Peak Area repeatability - < 1% RSD,
Column Oven:
Operating temperature range: +3 °C to 450 °C
Temperature set point resolution: 0.1 °C
Temperature stability 0.1°C or better
Maximum heating rate: 125 °C/min
Cryogenic option minimum temperature: Up to -50 °C
Number of ramps/plateaus: 32/33
Oven cool-down: 450 °C to 50 °C in less than 4 minutes.
The cooling methodology shall be automatic under processor control
For safety purpose oven power must turn-off automatically when the lid/door is opened.
Temperature programming & overheating protection.
Provision to install two or more general purpose columns
Possible to use capillary columns of 50, 100, 250, 320 microns and Above
Pneumatics: All Pneumatics Inlets must have an electronic pneumatic control (EPC) of carrier, split and septum
purge gases, which can compensate for variations in ambient temperature and pressure
for maximum stability including electronic ON/OFF
The pressure set points should be adjustable
by increments of 0.001 psi up to 100 psi.
with typical control of ± 0.001psi for the range of 0 to 150 psi
Maximum temperature attainable should be 400°c or more.
Flow & pressure set point for each inlet & detector
Carrier & makeup gases settings selectable for each gas, He, N2, H2
Inlet modules :
Pressure sensors ;
Accuracy < ± 2% full scale
Repeatability < ± 0.05 psi
Flow sensor
? Accuracy < ± 5% full scale
? Repeatability < ± 0.35 % of set point
Leak tests must provide preprogrammed leak tests available from monitoring software.
Inlets/Injectors/modes:
EPC compensated for atmospheric pressure & temperature variations
Instant Connect, user-exchangeable modules
Shall support different operation modes:
such as but not limited to split, splitless, split/splitless, pulsed split/splitless, large volume injections packed column simulation; with Head Space .
Maximum temperature: 400 Deg C or better
Suitable for all capillary columns (50 μm to 530 μm)
Electronic flow/pressure control with selectable range for better performance
? Total flow range able to set:
0 to 200 mL/min N2
0 to 1,250 mL/min H2 or He
Inlet sealing system for quick, easy, injector liner changes.
Programmable Temperature Injector
It should have facility to inject large volume through auto sampler
or manually up to 50 ul or more
Maximum temperature: 450 Deg C or better
Ramping rate up to 2000C/min or better.
Detectors: The system should have the following detectors:
? One Electron Capture Detector ECD detector with the following specification or better:
Minimum Detection Level: 50 fg /Sec or better.
Data Acquisition rate: up to 50 HZ.
Detector should reach: 400 °C temperature.
Linear dynamic range : 104 or better
Detector should have the facility to minimize the contamination and optimizes
sensitivity.
? One Flame Ionization detector FID detector with the following specification or better:
Minimum Detection Level: less than 2 pg N/s or better
Linear Dynamic range: > 107
Data acquisition rate: up to 500 Hz or better
All gases flow should be adjustable/controlled by software only
One Thermal Conductivity Detector TCD
Minimum Detection Level: less than 400 pg tridecane /mL or better
Linear Dynamic range: > 105 ± 5%
Nitrogen Phosphorus Detector NPD as optional detector with the following specifications
Minimum Detection Level: At least 0.1PG N /Sec or better.
Data Acquisition rate: up to 200 HZ or better.
Detector should reach 400 °C temperature.
All gases flow should be adjustable/ controlled by software with no manual control.
Auto sampler:
Liquid auto sampler for injection ports with optional manual operation.
It should inject from 0.1ul to 50.0ul with variable speed & varying syringe sizes & must be operate from software.
It should have capacity for 50 or more sample vials.
Programmable vial sequence.
It should have a reproducibility of <0.5% RSD.
Headspace sampler for volatile organic/samples:
Temperature range for Transfer line/ Sample up to 2000C or better.
Should have a closed sampling design without exposure to atmosphere or ambient temperature.
System must be syringe less to avoid exposure to atmosphere/contamination.
System must be controllable through software only.
PC / Software / Data processing:
PC with printer should be supplied along with the instrument.
Supplied PC compatible for operation & monitoring of GC.
High resolution LCD color display
8. Software related
Licensed software should be provided, Microsoft Windows operating platform with easy user interface
Basic software for device control, data analysis & supports a full range of printers
The GC must be able to interface to comprehensive real-time monitoring and diagnostic
software which includes all of the following items:
Real time notification via advisories and indicators,
Chromatographic attributes,
Instrument diagnostics, including leak tests,
Access to maintenance, run and event logs,
This system must be capable of creating and storing methods with stored parameters such as flow rate, pressure, column temperature
Report may be printed from all stored results format with results being combined from different runs and different measurement techniques.
It should be Possible to call these methods during a sequence of samples in the auto sampler.
The system should have the capability of running different methods for different samples.
All the methods, parameters should be through window-based software.
Software shall include quality control protocols.
Accessories:
The GC should be provided with necessary accessories. The presence of these accessories should not limit or restrict our usage as an independent GC.
All necessary gases with Gas Cylinders for Nitrogen, Hydrogen, Zero Air, and Helium, double stage Regulators, Gas Purification System for all gases, all necessary gas generators, connectors kit, column connectors and maintenance tool kit must be provided along with the system.
All necessary types of columns for training purpose.
Printer
Color LCD monitor
Consumables & spare parts:
All recommended Start up kit :-
Glass wares, Solvents for mobile phase , compatible vials, caps ,auto sampler Syringes, Ferrule (0.5 mm & 0.8 mm), Quartz Liners for Split/Split less mode, Quartz wool, Capillary adaptor, Pump Oil, Spare columns, spare tubing, fitting and tools for the maintenance of different aspects of the system
Hydrogen Generator Ferrules ,Septa Column and Nuts, Auto Sampler spares and vials, Injector & Detector heater & sensors ,Filters, Detectors & its spares, Universal Capillary Connectors, Injector Liners (for all makes), Graphite, Graphite/ Vespel,Teflon, Injectors (for All makes), Digital gas Flow meters
Other recommended parts & Consumables for two years operation. Considering expiry
???????????Power requirement: 220V, 50Hz for single phase, IEC World plug type F or EEC3/7 (Schuko) 380V, 50Hz for Three Phase
Warranty: 2 years starting from the date of acceptance testing approval
Installation and commissioning: The equipment is to be installed, configured, and thoroughly tested by the manufacturer before handing over
Training: Training of users in operation and basic maintenance to be provided for five days, 8 hours a day
Technical Documentation: User, installation and maintenance manuals is to be supplied in English language??????????????