Order Code: 22235922.4

Category: General Lab Equipment I

  System configurations (Lasers and colors) : minimum of 3-laser and12-color  Optical alignment : pre-aligned and fixed optical alignment system  Fluorescent resolution : ≤5% CV Flow cell : Fixed optical assembly with sp...



SPECIFICATION

 

  • System configurations (Lasers and colors) : minimum of 3-laser and12-color 
  • Optical alignment : pre-aligned and fixed optical alignment system 
  • Fluorescent resolution : ≤5% CV
  • Flow cell : Fixed optical assembly with spatially separated laser beams
  • Sample flow rates : minimum of 10 μL/min and up to 120 μL/min depending on low, medium and high flow rate 
  • Fluid capacity : Minimum of 4L tanks
  • Sample aspiration tubes, plates and tube : loader with minimum of 30-tubes (12 x 75-mm tubes), 96 well plates, and Manual tube port for 1.5ml and 2ml (micro centrifuge tubes),5ml (12 x 75-mm) polystyrene and polypropylene tubes, 15 and 50 ml falcon tubes (optional) 
  • Data Acquisition rate (Event rate) : Up to 30,000 events per second
  • Compensation : Automated or manual compensation 
  • System throughput : minimum 40 tubes multicolor assay per hour 
  • Threshold : single parameter or logical combination of multiple parameters
  • Data analysis software : integrated software to facilitate rapid and automated data analysis, bi-directional LIS interface
  • Peripheral ports : USB interface, Support for host computer, Support for external printers, Support for hand-held barcode reader
  • Peripheral Ports : USB interface, support for external printers, hand held barcode reader, Bi-directional LIS connectivity
  • Printer : Supplied with compatible colored external printer that uses cartilages available in the local market.
  • Barcode reader : in-built barcode reader and Support for hand-held barcode reader identification of sample and reagents
  • Operating Environment : Operating Temperature (+15?C to +40?C), relative humidity <85%
  • Utility requirements : Electrical power supply : 220VAC ±10%, Suitable UPS with maintenance free batteries for minimum one-hour back-up
  • Standards and safety Requirements : Shall meet IEC-60601 general electrical safety and shall meet ISO 13485 standards (or equivalent BSI)
  • Collection tubes : Supports different blood collection tubes (different size and multiple manufacturer’s) with minimum of 10 different tube types 
  • Reagent vials size and types : Allows use of multiple size reagent vials and load minimum of 30 reagents capacity, use of reagent vials from multiple manufacturers 
  • System can provides ready to acquire samples to the flow-cytometer
  • Bidirectional data integration with flow-cytometry 
  • Fluid tanks/reservoirs : include minimum of 1-10 L fluid tanks depending on type of fluid 
  • Dispensing volume : Sample (5–500 μL (in increments of 1 μL) and reagents(5–1000 μL (in increments of 1 μL)
  • Sample loading : minimum of 30 tubes at a time
  • Sample tube : use primary sample tube and secondary tubes 
  • Peripheral Ports : USB interface, support for external printers, Hand held barcode reader, Bi-directional LIS connectivity
  • Printer : Supplied with compatible external printer that uses cartilages available in the local market
  • Barcode reader : in-built barcode reader and Support for hand-held barcode reader identification of sample and reagents
  • Operating Environment : Operating Temperature (+15?C to +30?C), relative humidity <85%
  • Utility requirements : Electrical power supply: 220VAC ±10%, Suitable UPS with maintenance free batteries for minimum one-hour back-up
  • Standards and safety Requirements : Shall meet IEC-60601 general electrical safety and shall meet ISO 13485 standards (or equivalent BSI)
  • Documentation : User, technical and maintenance manual supplied in English
     