Earth Electrode Resistance Tester

Order Code: 22235966.2

Category: General Lab Equipment I

As per bs standards Ingress protection  IP54 C stake, P stake and Noise check Automatic Noise rejection 40 V pk to pk 2-wire, 3-wire, 4-wire test Yes, no shorting links required No disconnect  testing (ART) Yes, with MCC CLAMP Stakeless ...



SPECIFICATION

  • As per bs standards Ingress protection  IP54 C stake, P stake and Noise check Automatic Noise rejection 40 V pk to pk 2-wire, 3-wire, 4-wire test Yes, no shorting links required No disconnect  testing (ART) Yes, with MCC CLAMP Stakeless measurement Yes, with MCC CLAMP and MVC CLAMP Instrument output
  • Voltage: ±25 V or 150 V at 94 Hz, 105 Hz, 111 Hz and 128 Hz
  • Current: 4.5 mA or 0.45 mA or 0.045 mA Ground current range with clamp 0.5 mA to 19.9 A
  • Ground current accuracy 5% ±3 digits 
  • Ground voltage range 0 to 100 V ac 
  • Ground voltage accuracy 2% ±2 V 
  • Resistance ranges 2, 3, 4 pole: 0.01 Ω to 200 kΩ ART: 0.01 Ω to 200 kΩ 
  • Stakeless: 0.01 Ω to 200 Ω Resistance accuracy 2P measurements: 2% ±3 digits 3P measurements: 2% ±3 digits 4P measurements: 2% ±3 digits ART measurements: 5% ±3 digits Stakeless measurements: 7% ±3 digits
  • Maximum probe resistance Rp limit: 200 kΩ (50 V output voltage) Rc limit: 200 kΩ (50 V output voltage) Limits reduced to 10kΩ for 25 V output voltage Limits  reduced to 5 kΩ for 0.01 Ω resolution 
  • Display 31⁄ 2 digit high contrast liquid  crystal, backlit
  • Battery type DET4TC2: 8 off AA (LR6) cells DET4TCR2: 8 off  AA (LR6) NiMH rechargeable cells 
  • Operating temperature range  -15 °C to  +55 °C / 5 °F to 131 °F


 