CHANNEL - A  Range : 10Hz to 10MHz direct counter 10MHz to 100MHz perscaled by proportion  Resolution : Direct counter : 1, 10, 100Hz switch selectable : Per scaled : 10, 100, 1000Hz switch selectable 



SPECIFICATION

Frequency measurements
CHANNEL - A 

  • Range : 10Hz to 10MHz direct counter 10MHz to 100MHz perscaled by proportion 
  • Resolution : Direct counter : 1, 10, 100Hz
  • switch selectable : Per scaled : 10, 100, 1000Hz switch selectable 
  • Gate time : 0.01s, 0.1s is switch selectable 
  • Accuracy : ± 1 count ± time base error x measured 

CHANNEL - B 

  • Range 100MHz - 2.7GHz 
  • Resolution 100Hz, 1KHz, 10KHz 
  • Gate Time 0.01s, 0.1s, 1s 
  • Accuracy ± 1 count ± time base error x measured frequency 

Period Measurements 

  • Input : Channel A Range : 10Hz to 10MHz 
  • Resolution : 10-7 s, 10-8 s, 10-9 s, switch selectable 
  • Accuracy : ± 1 count ± time base error x measured preiod
  • Totalize measurements 
  • Input : Channel B 
  • Range : 10Hz to 10MHz 
  • Resolution : ± 1C count of Input 
  • Check : 8 Bits Repeating Display 0-9 Together 

CHANNEL - A 

  • Input Sensitivity 
  • : 10 Mhz range : 10Hz ~ 8MHz 20m Vrms 
  •  : 8MHz ~ 10MHz 30m Vrms 
  • : 100 Mhz Range : 10Hz ~ 8MHz 20m Vrms 
  •  : 8MHz ~ 10MHz 30m Vrms 
  • Attenuation : x 1, x 20fixed 
  • Filter (CHA only) Low pass : Ac 100KHz - 3dB 
  • Impedance : Approx, 1M Ohmless then 35pF 
  • Maximum voltage without damage : 250v (DC + Acrms) 

CHANNEL - B 

  • Input Sensitivity : 20m Vrms 
  • Impedance : Approx. 5 Ohmless 
  • Maximum voltage without damage : 3V 

Time Base

  • Time Base Frequency : 110MHz 
  • Short Term Stability : ± 3 x 10-9 for 1s 
  • Average Long Term Stability : 2 x 10-5 month 
  • Temperature : ± 1 x 10-5 0ºC to 40ºC 
  • Line Voltage : ± 1 x 10-7 for 10% change 

General

  • Display : 8 digit, 0.39 inch red led display with decimal point, gate, overflow, Khz, Mhz and ± mS indication 
  • Power requirement : Line 220V ± 10% 45hz ~ 70hz 
  • Warm-up Time : 20 minutes when cold started at 25ºC 
  • Temperature : Rated range of use : -5ºC ~ + 50ºC 
  • storage and transport : -40ºC ~ + 60ºC 
  • Humidity : Operating : 10 ~ 90 % RH 
  •  : Storage : 5 ~ 95%RH 
  • Dimension and weight : width : 207 mm 
  •  : height : 85 mm 
  •  : depth : 255 mm 
  •  : weight : 2 kg