Clinical Chemistry Analysers

Order Code: 22235922.99

Category: General Lab Equipment I

Minimum Test Menu/ Test parameters : Basic clinical chemistry Analytes, Electrolytes, Therapeutic drug monitoring(TDM),drug of abuse, proteins, etc  Specimen type: Serum, Plasma, whole blood, urine, CSF, and other body fluids  Throug...



SPECIFICATION

  • Minimum Test Menu/ Test parameters : Basic clinical chemistry Analytes, Electrolytes, Therapeutic drug monitoring(TDM),drug of abuse, proteins, etc 
  • Specimen type: Serum, Plasma, whole blood, urine, CSF, and other body fluids 
  • Throughput: Photometric part:-Minimum 300 tests per hour : Electrolyte part:- Minimum 150 samples per hour 
  • A minimum of Onboard Reagent Capacity : 40
  • Cuvette Washing : Automatic or disposable 
  • Measuring Technology: Photometric, Ion Selective and other standard technologies 
  • Clot Detection System: Equipped with sample, clot and bubble detection systems
  • Minimum Sample Loading Capacity: 75
  • STAT Option: Mandatory
  • Quality Control: minimum two levels
  • Bar Code Reader: Internal barcode reader for reagents and samples
  • System: closed system
  • Operating Mode: End point, Kinetic, initial rate, monochromatic,dichromatic, turbidimetric, serum blank (differential), fixed time, and other operating modes 
  • Wavelength Range: Automatic Wavelength selection by multi-position filter wheel – selection from wavelength of 340 – 800 nm.
  • Temperature Control: cuvette heating (electrical) in carousel and reading path: 37±0.2°C
  • System Interfaces: RS232 serial interface, bidirectional
  • UPS & Stabilizer : Each analyzer unit must be supplied with compatible UPS & stabilizer capable of supporting the instrument functions at times of line power outages for a minimum of one hour
  • Printer: Each analyzer unit must be supplied with a compatible printer that uses cartilages available in the local market 
  • Water System: Each analyzer unit must be supplied with a water purification system that can continuously feed the instrument with deionized and bacteria free water at all times of the instrument’s operation and with extra filters sufficient for at least one year’s consumption.
  • Operating Environment; Operating Temperature: +15 °C to +30°C, Relative humidity: < 85%
  • Utility requirements: Electrical Power Supply Requirement : 220VAC +10%
     