Automated Hematology Analyzer

Order Code: 22235922.98

Category: General Lab Equipment I

Test Range: Complete blood count (CBC), 5-part WBC differential, reticulocyte count, body fluid cell count Specimen type: Whole blood and other body fluids? Technology/methods :Flow-cytomery, electrical Impedance, spectrophotometry and other sta...



SPECIFICATION

  • Test Range: Complete blood count (CBC), 5-part WBC differential, reticulocyte count, body fluid cell count
  • Specimen type: Whole blood and other body fluids? Technology/methods :Flow-cytomery, electrical Impedance, spectrophotometry and other standard technologies 
  • Throughput:-A minimum of 80 samples per hour for CBC with differential. 
  • Aspirated Sample Volume and Type: Maximum of 200μl of whole blood and other body fluids 
  • Calibration and QC: Automatic calibration and three level quality controls
  • Operation Mode/Automation: Fully automated, autoloader capability , capability for flagging outlying results, pre-diluted blood analysis options
  • Tube Sampling: Both open and closed sampling options
  • STAT Options: STAT option is mandatory 
  • Result Analysis & Storage: Minimum of 10,000 patient results with graphs and patient information and interfaced with LIS system
  • Peripheral Ports: USB interface, support for external printers, hand-held barcode reader, Bi-directional LIS connectivity
  • Printer: Supplied with compatible external printer that uses cartilages available in the local market
  • Barcode reader: in-built barcode reader and Support for hand-held barcode reader
  • Operating Environment: Operating Temperature (+15?C to +30?C) 
  • Utility Requirement: Electrical power supply: 220VAC ±10%, UPS with maintenance free batteries for minimum one-hour back-up
  • Standards & safety Requirements: Shall meet IEC-60601 general electrical safety and shall meet ISO 13485 standards (or equivalent BSI

Enquiry Form