Order Code: 19204136.1.4
Category: General Lab Equipment III
Quantity Specifications 1 Measuring inputs 9 ALMEMO® input sockets, Electrically isolated with semiconductor relay (50 V) A/D converter Delta-si...
Quantity |
Specifications |
1 |
Measuring inputs 9 ALMEMO® input sockets, Electrically isolated with semiconductor relay (50 V) |
A/D converter Delta-sigma 24-bit, 50 mea. op. p. sec., electrically isolated. |
|
Sensor power supply -Rechargeable battery 9 or 12 V, maximum. 100 mA -Mains adapter 12 V, maximum. 100 mA |
|
Outputs 2 ALMEMO® sockets for all output modules (analog, data, trigger, relay cables, memory, etc.) |
|
Standard equipment -Display: Graphics, 128x128 pixels, 16 rows Lighting: 5 white LEDs, 3 levels -Keypad: 9 membrane keys (4 soft-keys), thumb-wheel -Memory: 512-KB EEPROM (100,000 meas. values) -Date and time-of-day: Real-time clock, buffered with battery |
|
Power supply -Rechargeable batt. pack 6 NMH rechargeable batteries, 1600 mA 2.5-hour high-speed charging, internal circuit - Mains adapter: 230V AC to 12 V DC, 1A, electrically isolated -DC adapter cable: electrically isolated 10 to 30 V, 1 A - Current consumption without input and output modules: Active mode approx. 37 mA with lighting approx. 45 to 100 mA. Sleep mode approx. 0.05 mA |
|
Housing (LxWxH) 204 x 109 x 44 mm ABS (maximum 70 °C), 550 g |
|
|
Calibration software AMR-Control software MA28909 Option KL :Multi-point calibration, special ranges |
|
ALMEMOManual, manufacturer’s test certificate, |
|
Calibration certificate of each parameter (atmospheric pressure, humidity and air temperature |
|
Mandatory accessories
Measuring range:700 to 1050mbar Overload:max.1.5 fold of fin.cal range Accuracy:±0.5% of final value Operating range:-10 to +600c;10 to 90%H non condensing Dimensions:90*20*7.6mm (MINIMUM) Hose connection:Ø5mm,12mm long Sensor material:aluminium Connecting cable:0.2m Extension cable:2m
Item name :Resistant thermometer Measuring element:Pt100 4L, DIN class B Options:DIN class A, 1/5 DIN class B Sensor tip: Diameter d1 Sensor tip: 3.2 mm Operating temp. Sensor tip: -40 to 500°C Length L: 250 mm, material: stainless steel Diameter d2, Sheathed line:2.8 mm Cable sleeve: Brass, hexagonal, L = 65 mm, circum diameter = 9 mm, operating temp. –40 to +160 °C Standard cable:1.5 meters line, FEP / silicone, insulated, operating temperature –50 to +200 °C Cable options: Line, PVC / PVC, insulated, operating temperature –20 to +105 °C
Humidity Measuring range 5 to 98 % RH Accuracy ±2.0 % RH in range 10 to 90 % at nominal temperature Hysteresis typical ±1 % RH Nominal temperature +23 °C ±5 °C Sensor operating pressure Atmospheric pressure Response time T63typical 8 seconds at +25 °C, 1 m/s(without filter)
Temperature Sensor NTC Type N Operative range -20 to +60 °C Accuracy typical ±0.2 K at 5 to 60 °C maximum ±0.4 K at 5 to 60 °C maximum ±0.7 K at -20 to +80 °C Reproducibility typical ±0.1 K Response time T63typical 20 seconds (without filter) Standard cable(connecting cable):1.5 meters line, FEP / silicone, insulated, operating temperature –50 to +200 °C |