AC/DC Current Probe

Order Code: 19204149.56

Category: General Lab Equipment III

Features ? Working temperature 23° C ±5° C ? 20% to 75% relative humitidy ? Probe zeroed ? 1 minute warm up ? Battery at 9V ± 0.1V ? External magnetic field < 40 A/m ? No adjacent current carrying conductor ? 1 M&Omeg...



SPECIFICATION

Features
? Working temperature 23° C ±5° C
? 20% to 75% relative humitidy
? Probe zeroed
? 1 minute warm up
? Battery at 9V ± 0.1V
? External magnetic field < 40 A/m
? No adjacent current carrying conductor
? 1 MΩ /100 pF probe termination

 

Measurement
Accuracy (Input Current:                                   100 mV/A                                     10 mV/A 
  dc to 1 kHz)                                                  Sensitivity Setting                      Sensitivity Setting
50 mA to 10A                                                                                                     < ±3% of reading ±50 mA    
50 mA to 40A                                                                                                     < ±4% of reading ±50 mA
40A to 100A                                                                                                       < ±15% of reading ±100 mA

Bandwidth: 

  • 3 dB Current measuring range: 
  • dc to 100 kHz Maximum Floating Voltage (refer to safety warnings and Standards)
  • 0A to 100A dc or peak ac
  • Phase shift: 
    1. < 1.5°dc to 65 Hz
    2. < 1.0°dc to 65 Hz
  • Noise:  3mV/ 480 μV
  • Slew Rate: 0.3 V/μs , 20 mV/μs
  • Insertion Impedance: 0.01Ω
  • Rise or Fall Time: 3 μs
  • Influence of Adjacent Conductor: <0.2 mA/A ac
  • Influence of Conductor Position in Jaw: 0.5% of reading in jaw
  • Overload Indication: Red LED (OL) indicates input is out of range
  • Maximum Working: 600Vac RMS CAT II
  • Voltage (refer to safety warnings and standards): 300Vac RMS CAT III
  • Maximum Floating: 600Vac RMS CAT II
  • Voltage (refer to safety warnings and standards): 300Vac RMS CAT III