Order Code: 22236235.49

Category: General Lab Equipment II

Isolation amplifier, channels A, B, C, E           Frequency range: dc to 80 kHz Input voltage (between 0 and U) Max 620 Vdc/460 Vac  Input resistance Ri = 1 MW in all ranges  Three?stage attenuato...



SPECIFICATION

  • Isolation amplifier, channels A, B, C, E          
  • Frequency range: dc to 80 kHz
  • Input voltage (between 0 and U)
  • Max 620 Vdc/460 Vac 
  • Input resistance Ri = 1 MW in all ranges 
  • Three?stage attenuator: MT = 1: 1; 1/10; 1/100 
  • Accuracy: ±2% of full scale range 
  • Input current (between 0 and I) 
  • Max: 10 A continuous; 16 A for t< 15 min; 20 A for t<2 min. 
  • Internal resistance: 30 mW in all ranges 
  • Two?stage attenuator: MT = 1 V/A; 1/3 V/A 
  • Accuracy: ±5% of full scale range 
  • Five outputs: A, B, C, D, E with led for over range indication 
  • Output resistance RO: 100 W 
  • Multiplexer: Mux channels, selectable: 1 to 8 (4 x signal; 4 x zero line) 
  • Gain attenuator, adjustable: 0.2 to 1
  • Y?position, adjustable: ?8 V to + 8 V
  • Trigger source, switchable to A, B, C, D, E
  • Mux frequency, adjustable: 50 kHz to 500 kHz (typical)
  • Two BNC outputs for oscilloscope 
  • Mathematical module and filter:  Functional modes for channel D: Addition A+B; subtraction A?B; multiplication AxB/10 or AxB; reconstruction of the phase  voltage LIN(A, B, C) from the line?to line voltages; channel E  switched into channel D for multiplexing
  • Filter: Low pass active filter of the 2° order required for the recovery  of the fundamental wave out of the PWM signals. 
  • Cut?off frequency: 1 kHz
  • Space vector indicator       
  • Voltage vector: indication with 7 led
  • Magnetic flux vector: BNC outputs X e Y for oscilloscope
  • Power supply: Single?phase from mains Frequency: 50/60 Hz
     