Insulation Tester (Battery Operated)

Order Code: 19203351.13

Category: General Lab Equipment III

 FEATURES  Bargraph to display insulation resistance.  Displays the value of external AC voltage along with flashing symbol.  Auto null function to automatically subtract the test lead resistance before displaying the re...



SPECIFICATION

 FEATURES

  1.  Bargraph to display insulation resistance.
  2.  Displays the value of external AC voltage along with flashing symbol.
  3.  Auto null function to automatically subtract the test lead resistance before displaying the real continuity resistance value.
  4.  Live circuit warning beeper.
  5.  Releasing the test button automatically discharges the charges stored in the circuit under test.
  6.  200mA continuity measuring current to IEC 61557.
  7.  Minimum 1mA current on insulation tests to IEC 61557.

 SPECIFICATION

  1.  INSULATION RESISTANCE
    1.  Test voltage : 250V/500V/1000V
    2.  Measuring ranges (Mid-scale value) : 20M/200M/2000M
    3.  Output voltage on open circuit : Rated test voltage+20%,
    4. 0% - Nominal current : 1mA DC min.
    5.  Output short circuit current : 1.5 mA DC approx.
    6.  Accuracy : ±1.5%rdg±5dgt (20M/200M), ±10%rdg±3dgt (2000M)
  2. CONTINUITY TEST
    1. - Measuring ranges : (Mid-scale value) 20/200/2000
    2. - Output voltage on open circuit : 7
    3. - 12V DC - Measuring current : 200mA DC min.
    4. - Accuracy : ±1.5%rdg±5dgt (20), ±1.5%rdg±3dgt (200/2000)
  3.  AC VOLTAGE
    1.  AC voltage range : 0
    2.  600V AC - Accuracy : ±5%rdg±3dgt
  4. GENERAL
    1. Safety Standard : IEC 61010-1 CAT III 300V Pollution Degree 2, IEC 61010-2-031
    2.  IEC 61557-1/2/4
    3.  IP Rating : IEC 60529-IP54
    4.  EMC Standard : IEC 61326-1
    5.  Power source : R6 (AA) (1.5V) × 8

 ACCESORRIES - Test leads Pouch for test leads - Instruction Manual
 MANUAL This equipment must come with complete user's instruction manual in English.